R3132 Series Spectrum Analyzer Operation Manual
5.2.20 Sweep Time Accuracy
5-85
Table 5-26 Sweep Time Accuracy Test Setting
For R3132 series For signal generator
Sweep Time Pulse Period Pulse Width
50 sec 45 sec 1 sec
5 sec 4.5 sec 1 sec
500 msec 450 msec 200 msec
50 msec 45 msec 20 msec
5 msec (OPT29) 4.5 msec 2 msec
500 µsec (OPT29) 450 µsec 200 µsec
50 µsec (OPT29) 45 µsec 20 µsec