R3132 Series Spectrum Analyzer Operation Manual
5.2.7 IF Gain Uncertainty
5-28
Table 5-10 IF Gain Uncertainty Measurement Setting
1 dB Step Attenuator 10 dB Step Attenuator Reference Level
1 dB 0 dB -1 dBm
2 dB 0 dB -2 dBm
3 dB 0 dB -3 dBm
4 dB 0 dB -4 dBm
5 dB 0 dB -5 dBm
6 dB 0 dB -6 dBm
7 dB 0 dB -7 dBm
8 dB 0 dB -8 dBm
9 dB 0 dB -9 dBm
0 dB 10 dB -10 dBm
0 dB 20 dB -20 dBm
0 dB 30 dB -30 dBm
0 dB 40 dB -40 dBm
0 dB 50 dB -50 dBm

Table of Contents