R3132 Series Spectrum Analyzer Operation Manual
5.2.14 Accuracy of Frequency Readout and Count Frequency Marker
5-54
Table 5-18 Frequency Readout Accuracy Test Setting
Table 5-19 Count Frequency Marker Accuracy Test Setting
Signal Generator Center Frequency Frequency Span Note
2 GHz 2 GHz 1 MHz
2 GHz 2 GHz 10 MHz
2 GHz 2 GHz 20 MHz
2 GHz 2 GHz 100 MHz
2 GHz 2 GHz 1000 MHz
5 GHz 5 GHz 1 MHz For R3162/72/82
5 GHz 5 GHz 10 MHz For R3162/72/82
5 GHz 5 GHz 100 MHz For R3162/72/82
5 GHz 5 GHz 1000 MHz For R3162/72/82
11 GHz 11 GHz 1 MHz For R3172/82
11 GHz 11 GHz 10 MHz For R3172/82
11 GHz 11 GHz 20 MHz For R3172/82
11 GHz 11 GHz 100 MHz For R3172/82
11 GHz 11 GHz 1000 MHz For R3172/82
18 GHz 18 GHz 1 MHz For R3172/82
18 GHz 18 GHz 10 MHz For R3172/82
18 GHz 18 GHz 20 MHz For R3172/82
18 GHz 18 GHz 100 MHz For R3172/82
18 GHz 18 GHz 1000 MHz For R3172/82
35 GHz 35 GHz 1 MHz For R3182
35 GHz 35 GHz 10 MHz For R3182
35 GHz 35 GHz 20 MHz For R3182
35 GHz 35 GHz 100 MHz For R3182
35 GHz 35 GHz 1000 MHz For R3182
Signal Generator Center Frequency Note
2 GHz 2 GHz
5 GHz 5 GHz For R3162/72/82
11 GHz 11 GHz For R3172/82
18 GHz 18 GHz For R3172/82
35 GHz 35 GHz For R3182