R3132 Series Spectrum Analyzer Operation Manual
3.3.26 SWEEP Key (Sweep Time)
3-75
Gate Position Sets the position of a gate signal.
Gate Width Sets the width of a gate signal.
Gated Sweep ON/OFF Turns the gated sweep mode on or off.
ON: Sweep is executed according to the currently set gate
conditions (gate position and width).
OFF: Turns the gated sweep mode off.
Screen A/B Toggles the active screen between A and B.
A: The upper screen is set to the active state.
B: The lower screen is set to the active state.
Gate Setup Quit The trigger condition setup screen of gate signals is released to re-
turn to the Sweep menu.
Gated Sweep ON/OFF Turns the gated sweep mode on or off.
ON: Sweep is executed according to the currently set gate
conditions (gate position and width).
OFF: Turns the gated sweep mode off.
Ext Gate In ON/OFF Changes the input signal at the External Trigger connector.
ON: Performs a gated sweep using the External Trigger con-
nector signal as the gate signal.
Sweeps intermittently when the signal level is greater
than approximately 2.5V, and stops sweeping when the
signal level is approximately less than 2.5V.
OFF: Performs a normal sweep.
Uses the input signal at the External Trigger connector
as the trigger signal for the Ext Trig mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3132 Series and is the answer not in the manual?

ADVANTEST R3132 Series Specifications

General IconGeneral
BrandADVANTEST
ModelR3132 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals