EasyManuals Logo

ADVANTEST R3132 Series User Manual

ADVANTEST R3132 Series
588 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #355 background imageLoading...
Page #355 background image
R3132 Series Spectrum Analyzer Operation Manual
5.2.11 Residual FM
5-44
23. On the R3132 series, press as follows to capture minimum peak signal.
MKR, Peak Menu, Min Peak
24. Record the level of delta marker reading as Delta Level on the performance ver-
ification record sheet.
Calculation Residual FM
25. Calculate the Residual FM using the following formula.
Residual FM [Hz] = Slope [Hz/dB] × Delta level [dB]
Record the result on the performance verification record sheet.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3132 Series and is the answer not in the manual?

ADVANTEST R3132 Series Specifications

General IconGeneral
BrandADVANTEST
ModelR3132 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals