R3132 Series Spectrum Analyzer Operation Manual
5.2.16 Frequency Response
5-67
79. On the power meter, set the correction frequency to 30 GHz.
80. On the R3182, press as follows to tune the preselector.
FREQ, 1/2_more, Presel, Auto Tune
81. After the auto tuning has completed, adjust signal generator output level so that
the marker reading is -10 dBm ±0.09 dB.
82. Record the display of power meter reading with reverse sign in the performance
verification record sheet.
83. On the R3182, press FREQ and ∆ to increase center frequency by 200 MHz.
84. On the signal generator, increase the frequency of output by 200 MHz.
85. On the power meter, increase the correction frequency by 200 MHz.
86. Repeat steps 80 through 85 for every center frequency by 200 MHz step up the
center frequency to 40 GHz listed in the performance verification record sheet.