R3132 Series Spectrum Analyzer Operation Manual
5.3.5 Non Harmonic Distortion
5-106
2. On the R3132 series, after preset, set controls as follows:
Center Frequency: 0 Hz
Span: Zero
RBW: 10 MHz
TG: ON
TG Output Level: -10 dBm
3. On the spectrum analyzer, after preset, set controls as follow:
Stop Frequency: 6 GHz
Measuring the Non Harmonic Distortion
4. On the R3132 series, press ∆ key, to set center frequency by 10 MHz step up to
3 GHz.
5. Capture the biggest distortion signal except second harmonic signal.
6. Record the level difference between fundamental signal and biggest distortion
signal on the performance verification record sheet.