R3132 Series Spectrum Analyzer Operation Manual
5.3.6 TG Leakage
5-108
2. On the R3132 series, after preset, set controls as follows:
Start Frequency: 30 MHz
Stop Frequency: 3.0 GHz
Sweep Time: 20 sec
RBW: 1 kHz
VBW: 10 Hz
Reference Level: -60 dBm
TG: ON
TG Output Level: 0 dBm
(Ignore UNCAL message under above setting condition)
Measuring the TG Leakage
3. On the R3132 series, press SINGLE for single sweep.
4. On the R3132 series, after single sweep has completed, press PK SRCH for cap-
ture the peak signal.
5. Record the measurement data on the performance verification record sheet.