R3132 Series Spectrum Analyzer Operation Manual
6.1.1 External Mixer Performance Verification Test Procedures
6-7
11. On the R3172/R3182, press keys as follows to preset.
SHIFT, CONFIG (PRESET)
12. On the R3172/R3182, set controls as follows;
Center Frequency: 26.5 GHz
Span: 40MHz
Reference Level: 0dBm
dB/div: 10dB/div
RBW: 300kHz
VBW: 3kHz
Detector: Sample
13. On the SG, set controls as follows;
Output Frequency: 26.5 GHz
Output Level: 0 dBm
14. On the R3172/3182, press SINGLE key for a sweep.
15. On the R3172/R3182, press PKSRCH key to capture the signal peak.
16. Record the level of peak marker in the measured value column on the perfor-
mance verification test data sheets.
17. Calculate actual value using following equation.
Actual Value = Abs (Calibrated value) - Abs (Measured value)
18. Record the result in the actual value column on the performance verification test
data sheets.
19. On the R3172/R3182, set control as follow;
Center Frequency: 26.6 GHz
20. On the SG, set control as follow;
Output Frequency: 26.6 GHz
21. On the R3172/3182, press SINGLE key for a sweep.
22. On the R3172/R3182, press PKSRCH key to capture the signal peak.
23. Record the level of peak marker in the measured value column on the perfor-
mance verification test data sheets.
24. Calculate actual value using following equation.
Actual Value = Abs (Calibrated value) - Abs (Measured value)