R3132 Series Spectrum Analyzer Operation Manual
2.2.12 Intermodulation
2-47
Power on
2. Turn the power on.
Setting the signal generators
This prepares the signal generator outputs.
3. Set SG1 as follows: the frequency to 200.0 MHz; level to -4 dBm; modulation to
non-modulated mode; and output to ON.
4. Set SG2 as follows: the frequency to 200.2 MHz; level to -4 dBm; modulation to
non-modulated mode; and output to ON.
Each signal has an input level of -10 dBm.
Initialization
This resets the current settings to the factory defaults.
5. Press SHIFT and CONFIG(PRESET).
The default settings have now been reset.
Setting the measurement conditions
This changes the analyzer settings so that the input signal is displayed more clearly.
6. Press FREQ, 2, 0, 0 and MHz.
A center frequency of 200 MHz is set.
7. Press SPAN, 1 and MHz.
A frequency span of 1 MHz is set.
8. Press BW, RBW AUTO/MNL(MNL), 3 and kHz.
The RBW is set to 3 kHz.
9. Press VBW AUTO/MNL(MNL), 3, 0, 0 and Hz.
The VBW is set to 300Hz.
The attenuator level default setting is 10 dB, which makes the input to the mixer
-20 dBm (= -10 dBm - 10 dBm). Since the mixer level exceeds the distortion
limit, spurious peaks (3 and 4) appear in addition to the normal peaks (1 and 2).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3132 Series and is the answer not in the manual?

ADVANTEST R3132 Series Specifications

General IconGeneral
BrandADVANTEST
ModelR3132 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals