6.Alarm and troubleshooting
93
Table 6-1
Priority
Definition
Alarm Tone
Mute
Notice Message
High
Death or
irreversible
injury.
10 alarm tones
including 2hurry
alarm cycle is 8s.
120s
Red background,”!!!”flickering,
displaying frequency is 2Hz.
Medium
Reversible
injury.
3 alarm tones,
alarm cycle is 10s.
120s
Yellow background,”!!” flickering,
displaying frequency is 0.5Hz.
Low
Slight injury or
uncomfortable.
2 alarm tones, no
repeat.
120s
Cyan background,”!” flickering,
long-term display.
6.2. Alarm message
The alarm method includes technical alarm and functional alarm; technical alarm includes startup
test list and alarm occurring during normal operation.
6.2.1. Alarm default value and setting range
Table 6-2
Monitoring
parameters
High
Low
Setting range of high
limit
Setting range of
low limit
MVAdult
12
1
125 L/min
020 L/min
MVChild
6
1
PAWAdult
50
10
10110 cmH
2
O
070 cmH
2
O
PAWChild
40
8
FreqAdult
18
---
860 bpm
---
FreqChild
30
---
FiO
2
100%
18%
21%100%
18%99%
Hal.
1.5
OFF
0.1%8.4%
OFF,0.1%8.3%
Iso.
2.3
OFF
0.1%8.4%
OFF,
0.1%8.3%
Enf.
3.4
OFF
0.1%9.9%
OFF,
0.1%9.8%