8.Regular maintenance
129
No
Replaceable part
Every 6 month
Every 12 month
Every 36
month
Part Number
6.
Check /Replace
U-type sealing ring
No
Check
Replace
130004091
7.
Check /Replace
AGSS transmitted
pipe
No
Check
Replace
122001806
8.
Check /Replace
valve sealing
gasket
No
Check
Replace
130004147
9.
Check /Replace
O-ring
sealing(Pressure
gauge
Check
Replace
Replace
304000096
10.
Check /Replace
O-ring 14×2.65on
vaporizer mount
Check
Replace
Replace
304000044
11.
Sealing ring
1(Bypass circuit)
Check
Check
Replace
130013721
12.
Sealing ring
3(Bypass circuit)
Check
Check
Replace
130013723
13.
Clasp sealing ring
(Bypass circuit)
Check
Check
Replace
130013724
14.
Soda lime cans
Check
Check
Replace
130013683
15.
Sealing ring
(Bypass circuit)
Check
Check
Replace
130013689
16.
Sealing ring
(Bypass circuit)
Check
Check
Replace
130013690
CAUTION: The life-span of all components above is the result in normal circumstance
and work condition.
8.4. Maintenance of O
2
sensor
WARNING: Exhausted and old O2 sensor must be treated according to the present
local regulation. Not burn or dispose the O2 sensor casually.