CAUTION: According to this schedule to do maintenance timely and replace
corresponding parts.
CAUTION: The recommended repair time interval is 5 years. Breathing system,
vaporizer and parts are recommended to have a comprehensive test and replacement
every 5 year.
WARNING: The ineffective sodium lime has an obvious change in color when it’s
ineffective.
8.2.1. Appearance checklist
1. Assure the appearance of Aeon8800A no damage, the Anaesthetic Workstation can run
normally.
2. Assure absorbent canister install correctly and fill enough soda lime. Assure enough
anaesthetic in vaporizer.
3. Assure the exhaust gas absorber pipe connected with APL valve no damage and dry inside.
4. Assure O
2
, N
2
O, AIR, VAC and EVAC connection pipe no damage.
5. Assure O
2
, N
2
O, AIR, VAC and EVAC flexible pipe no damage (if they are be used).
6. Assure AC power cable no damage.
8.3. Replacement of replaceable parts