40
3.2.9. Electronic control flowmeter area
Name description
Diagram
The electronic control flow meter displays real-time O
2
and
N
2
O or O
2
and AIR flow values. Among them, O
2
is fixed
display, N
2
O and O
2
switch directly by clicking the button.
The electronic control flow meter guarantees a minimum O
2
concentration of 25% in an O
2
/N
2
O mixture. N
2
O cut-off if O
2
pressure is less than 200kPa.
Every flowmeter has 2 virtual flow tubes that display flow.
Each flow tube has the same icon, scale, structure and color
(grey). The units and numerical readings are displayed in
white. The fill color of virtual flow tubes is the same with gas
icon color.
Click on the button at the bottom of the flowmeter to adjust the
flow.
The flow meter can be adjusted and controlled in two ways
1. Adjust the flow of each gas;
2. Adjust total flow, and the ratio of one gas.