9.8.2. Ventilation parameter
Table 9-11 ventilator parameter
Item
Range
Stepping
Default Value
Remark
V
t
20mL~1500mL
10mL~1500mLoption
5(<100)
10(100~1000)
50(> 1000)
120mLChild;
500mLAdult
Have this function in IPPV SIMV~VC SIMV~VGPCV~VG mode
10mL~100mL, error is ±15% or ±10mLchoose the maximum one ;
over 100mLincluding 100 mL),error is ±15%
Freq.
2~60bpm(PS/CPAP );
2~100bpm(SIMV~VC
SIMV~PCSIMV~VG)
4~100bpm(Other modes,
excluding BIVENT;APRV)
1bpm
20bpmChild;
12bpmAdult
Error is ±1bpm or ±5%, choose the maximum one
I:E
4:1~1:8
0.5
1:2
Have this function in IPPVPCVPCV~VG mode,error is ±15%
T
INSP
0.2s~5.0s
0.1sec
1secChild;
2secAdult
Have this function in SIMV~PCSIMV~VCSIMV~VG mode, error is
±10% or 0.1s, choose the maximum one
PEEP
OFF, 3cmH
2
O~50cmH
2
O
1cmH
2
O
OFF
When it is OFF, PEEP value is 1cmH
2
O~3cmH
2
O; when
3cmH
2
O~50cmH
2
O, error is ±2cmH
2
O or ±15%, choose the maximum
one
Freq
MIN
2bpm ~60bpm
1bpm
4bpmChild;
2bpmAdult
Have this function in PS mode
T
P
OFF, 5%~60%
5%
10%
Have this function in IPPV SIMV mode, error is ±30% of the setting or
±0.05s, choose the maximum one