WARNING: Components must be cleaned before disinfection or sterilization.
7.3.3. Sterilization method
WARNING: The machine adopting autoclaving damp-heat mode must use deionized
water to avoid scale in the breathing system blocking the pipeline or entering the
patient's breathing tubes.
WARNING: Among plastic parts, only those marked with 134°C signs are
pressure-resistant and heat-resistant parts and can be sterilized by pressure steam at
134°C.
7.3.3.1. Pressure steam sterilization
Pressure steam sterilization should be the first choice for instruments, utensils, and articles that
are resistant to humidity and heat.
Sterilizer: Pulsation Vacuum Steam Sterilizer
Model: BIST-A-D360-D-A
Manufacturer: Shandong SHINVA Medical Devices Co., Ltd.
Temperature sensor: Wireless temperature recorder
Model: TL-2
Manufacture: Shantou DVP Technology Co. Ltd.
Table 7-3 Sterilization cycle
7.3.4. List of cleaning, disinfection, and sterilization methods
Table 7-4 Cleaning, disinfection, and sterilization list