118
7.4.1.2. Assembly
Connect the bellows to the breathing system by reassembling all components in the reverse order.
7.4.2. Disassemble the oxygen sensor
7.4.2.1. Disassembly
Disconnect one end of the O
2
sensor cable from the O
2
sensor socket.
Hold O
2
sensor, and unplug it from the inspiratory valve
cover with force.
Unscrew O
2
sensor connector from O
2
sensor and put it on
a safety place.

Table of Contents