9.Technical data
141
9.6. Input and output
NOTE: When the Anaesthetic Workstation stops transferring, the pipeline supplying
pressure should be 280kPa to 600kPa.
NOTE: When main power fails, the system switches to battery power automatically, and
all the Anaesthetic Workstation components can work normally.
NOTE: When main power and internal battery both fails, the gas output is not affected.
9.6.1. Electrical
WARNING: IEC60601-1 is applicable to the connections between medical devices, and
to the connections between at least one medical device and one or more non-electrical
devices. Even if no functional connections between the single devices, when using one
auxiliary power outlet, they form a medical electrical system. User must be aware of
the risk of increased leakage current and temperature and other increased risk may
exist, when the device is connected to the auxiliary socket-outlet.
WARNING: When earth wire has failure, if the device connects with auxiliary power
supply, it may cause patient current leakage exceeding the permitted range.
WARNING: If the exterior protective grounded wire has questions during wiring,
interior power supply must be used for device running.
Table 9-5
Supply Voltage
100-240VAC, 50/60Hz
Maximum Input Current
8A
AC power fuse
F10AH 250VAC, φ5X20F
Auxiliary AC power plug
4
Auxiliary AC power plug fuse
T2AH 250VAC, φ5X20F
Maximum output valve of auxiliary
AC power plug
1.5A(single plug); 6A(in total)
Grounded resistance
Small than 0.2Ω