1.Product description
1
1. Product description
1.1. Safety information
1.1.1. Warnings
Please read and adhere to all the WARNINGs, CAUTIONs, and NOTICEs listed below and
attached in their corresponding chapters throughout this manual.
WARNING: Statement gives important information that, if ignored, could lead directly
to personal injury and death.
WARNING: Statement gives important information that, if ignored, could lead directly
to damage to equipment and other property.
WARNING: Aeon8800A must only be operated by trained, skilled medical staff.
WARNING: Aeon8800A Anaesthetic Workstation must and can only be serviced by
Aeonmed’s authorized service representatives.
WARNING: For a full understanding of the performance characteristics of this
equipment, the user should carefully read this manual before operating:
All system connections.
All WARNINGs and CAUTIONs.
How to use the Anaesthetic Workstation.
How to test the Anaesthetic Workstation.
WARNING: Do the following things before operating Aeon8800A:
Finish all the tests listed in Pre-use Preparation.
Test other systematic parts.
WARNING: Do not use the machine if any test failures. Contact Aeonmed’s authorized
service representative for maintaining the machine.
WARNING: The Aeon8800A is not suitable for use in a magnetic resonance imaging
MRIenvironment.
WARNING: To avoid the risk of electric shock, this equipment must only connect to a
supply main with protective earth.
WARNING: The highest attitude of Aeon8800A Anaesthetic Workstation is 3000m.
WARNING: The pipeline inlet pressure of the Aeon8800A Anaesthetic Workstation must
be within 280-600kPa.
WARNING: Breathing tube, mask, SUCTION filter, adapter of mainstream CO
2
module,
etc. are disposable accessories. Check the state of the above accessories before use
to avoid repeated use.
WARNING: When the suction flow is insufficient, the suction port filter should be
checked or replaced.
WARNING: Do not use anti-static and/or electric breathing tubes or masks; otherwise,
they will be flammable near high frequent surgery equipment.