2.Components
25
2.6. Negative pressure suction device (optional)
An optional suction device is available to remove secretions, blood or other semi-liquid fluid from
the patient’s respiratory tract.
Before performing suction, adjust the negative pressure on the Suction device to an appropriate
value. When using the Suction device, connect the gas inlet to the Aeon8800A auxiliary gas
supply port. If the Suction device is connected with the other gas supplies, the pressure of the gas
supply should be less than 0.4MPa.
If suction flow is insufficient, check the suction port filter.
Table 2-8 Negative pressure suction device
No
Name
No
Name
1
Negative pressure control switch
4
Overfill protection device
2
Negative pressure indicator
5
Negative pressure connector
3
Negative pressure control
6
Positive pressure input port
CAUTION: The negative pressure suction device used together with the Anaesthetic
Workstation conforms to EN ISO 10079-3.
CAUTION: For more information about operation and maintenance of the negative
pressure suction device, please refer to the Instruction for Use provided with the
negative pressure suction device.