6.Alarm and troubleshooting
107
Paramagnetic oxygen
communication failure
Paramagnetic oxygen
connection problem or
paramagnetic oxygen fault
Contact the manufacturer for
maintenance.
Paramagnetic oxygen
failure (optional)
Paramagnetic oxygen
damage
Contact the manufacturer for
maintenance
Paramagnetic oxygen
pump not connected
Paramagnetic oxygen pump
connection problem.
Contact the manufacturer for
maintenance
Paramagnetic oxygen
pump failure
Paramagnetic oxygen pump
failure
Contact the manufacturer for
maintenance
Paramagnetic oxygen
sampling blockage
The paramagnetic oxygen
sampling tube is blocked.
Contact the manufacturer for
maintenance
Solenoid valve
(paramagnetic oxygen)
failure
Contact the manufacturer for
maintenance
6.3. Failure diagnosis
Table 6-5
The connection or
setting of APL valve is
incorrect.
1. Check the installation and connection
of APL valve;
2. Check whether the exhaust port is
blocked or not.
High breathing
system
temperature
Breathing system is
too high.
1.Disconnect the breathing system
heating cable and call service Rep.
Gas monitoring
has no sampling
signal
Sample line is blocked
or unconnected.
1. Check sample line and Y-piece filter.
2. Replace if necessary.
Incorrect connection of
pipeline or line
1. Leakage from incorrect connection of
the breathing tubes.
2. Incorrect installation of absorber
canister.
3. Incorrect installation of Bellows and
Bellows cover,
4. Incorrect installation of O
2
sensor or O
2