104
NO
Alarm Name
Priority
Alarm
type
Cause
Remedy
failure
failure.
60.
Paramagnetic
oxygen
sampling
blockage
High
Technical
alarm
The
paramagnetic
oxygen sampling
tube is blocked.
Check the sampling tube
and patient end filter.
61.
FSU High
pressure
High
Function
alarm
The Airway
pressure
exceeds the
upper alarm limit.
The applied
suction pressure
is higher than the
set value.
Check the ventilation
settings.
Check the alarm limits.
The breathing
hose is blocked
or the Airway
intubation is
kinked.
Check the respiratory circuit
and endotracheal intubation.
6.2.3. Self-test alarm
Table 6-4
NO
Alarm name
Alarm reason
Maintenance measures
1
Oxygen source sensor
failure
Oxygen source sensor
failure
Contact the manufacturer for
maintenance
2
Air source sensor failure
Air source sensor failure
Contact the manufacturer for
maintenance
3
Oxygen source failure
The oxygen source is not
connected
Check the oxygen supply.
4
Air source failure
The Air source is not
connected
Check the Air supply.
5
Failure of oxygen flow
meter
Oxygen flow meter failure
Contact the manufacturer for
maintenance
6
Air flow meter failure
Air flow meter failure
Contact the manufacturer for
maintenance
7
Nitrous oxide flowmeter
failure
Nitrous oxide flowmeter
failure
Contact the manufacturer for
maintenance