6.Alarm and troubleshooting
103
NO
Alarm Name
Priority
Alarm
type
Cause
Remedy
filter.
52.
Gas module
sample line
blocked
Low
Technical
alarm
The gas module
sampling pipe is
blocked.
Check the sampling tube
and patient end filter.
53.
Low battery
Middle
Technical
alarm
The battery
charge is less
than 10%.
Restore AC power.
54.
Network power
failure
Low
Technical
alarm
The device is not
connected to AC.
Restore AC power.
55.
The oxygen flow
does not reach
the set value
Low
Function
alarm
The oxygen
source pressure
is lower than
200KPa
Check the oxygen supply.
Internal pipeline
leakage
Please contact the
manufacturer for
maintenance.
56.
N2O flow does
not reach the
set value
Low
Function
alarm
Nitrous oxide
source pressure
is lower than
200kPa
Check the Air supply.
Internal pipeline
leakage
Please contact the
manufacturer for
maintenance.
57.
The Air flow
does not reach
the set value
Low
Function
alarm
Air source
pressure below
200kPa
Check the Air supply.
Internal pipeline
leakage
Please contact the
manufacturer for
maintenance.
58.
Paramagnetic
oxygen pump
not connected
High
Technical
alarm
There is a
problem with the
connection of
paramagnetic
oxygen pump.
Contact the manufacturer
for maintenance.
59.
Paramagnetic
oxygen pump
High
Technical
alarm
Paramagnetic
oxygen pump
Contact the manufacturer
for maintenance