1.Product description
3
WARNING: Disconnect the power plug from the main supply before removing the rear
panels or servicing the machine.
WARNING: Independent means of ventilation (e.g. a self-inflating manually powered
resuscitator with mask) must be available whenever the machine is in use.
WARNING: A malfunction of the central gas supply system may stop the operations of
more than one device that connects to the system simultaneously.
WARNING: Use cleaning and sterilizing schedule that confirms your institution’s
sterilization and risk-management policies.
Refer to the material safety data as applicable.
Refer to the operation and maintenance manuals of all sterilizing equipment.
A damaged oxygen sensor may lead to leakage of its highly caustic contents that
contain potassium chloride. Wear safety gloves and safety eyeglasses per the
manufacturers recommendations.
Do not inhale fumes that may result from any sterilization process.
WARNING: Use extreme care while handling the absorbent as it is a caustic irritant.
WARNING: Use care in lifting and manipulating vaporizers during the mounting
process as their weight may be greater than expected, based on their size and shape.
WARNING: All gas supplies should be of medical grade.
WARNING: To avoid endangering a patient, do not perform testing or maintenance
when the machine is in use.
WARNING: All gas volume, flow, and leakage specifications have been tested under
STPD, except those associated with the Anaesthetic Breathing system, which is tested
under BTPS.
WARNING: After the device replacement is finished, all calibrations must be done
again and only the service engineer is allowed to do this.
WARNING: If the Anaesthetic Workstation is equipped with Anaesthetic Gas Delivery
System, Monitoring Equipment, Alarm System, and Production Device, no matter
whether the machine is supplied by a separate device or system assembly, the
Anaesthetic Workstation checklist must be provided.
WARNING: The medical device connected to the Anaesthetic Workstation must
conform to IEC 60601-1 and IEC 60601-1-2
1.1.2. Cautions
CAUTION: For your safety and that of your patients, strictly follow this user manual.
CAUTION: Perform the tests specified on the Pre-use Preparation and, in case of a fault,
do not operate the machine until the fault has been corrected.
CAUTION: Before starting the machine, users must be familiar with the information
contained in the User Manual.