WARNING: Do not incline the machine at an angle of more than 10 degrees.
WARNING: Perform the pre-use checkout before operating the Aeon8800A.
WARNING: Remove all equipment on the top cover before moving the machine. Use
the handle on the machine to move the machine.
WARNING: Ensure that the machine does not incline when it is upgrading,
downgrading, turning around, or getting across a threshold. Do not push the machine
across pipelines, lines, or any barrier on the ground.
WARNING: Remove all the equipment attached to the flank of the machine before
transportation, otherwise, the incline of the machine will lead to personnel damage.
WARNING: Lock the castor wheel when the Anaesthetic Workstation is in use.
WARNING: Do not put the item over 25 kg on the top board of the Aeon8800A
Anaesthetic Workstation.
WARNING: Do not put the item over 12 kg on the work table of the Aeon8800A
Anaesthetic Workstation.
WARNING: The whole width of the machine should be less than 0.8m when hanging
some stuff.
WARNING: To protect patients and users from electrical hazards, it is imperative that
all systems consisting of electrical medical devices and other electrical devices, such
as but not limited to PCs, printers, etc., be mounted exclusively by trained personnel.
WARNING: The medical gas pipeline system malfunction could cause one or more
devices connected to the system to stop their operation; this does not apply to the
Anaesthetic Workstation that only uses cylinders for gas supply.
WARNING: The parts of the machine in contact with gas to be inhaled by patients don’t
contain phthalates, which are known to be carcinogenic, mutagenic, or toxic to
reproduction.
WARNING: Possible explosion hazard. Do not operate the machine near flammable
anaesthetic agents or other flammable substances. Do not use flammable anaesthetic
agents (i.e., either cyclopropane).
WARNING: Only non-flammable anaesthetics such as desflurane, sevoflurane,
enflurane, and isoflurane are allowed in the anaesthetic system, for they are
compatible with fresh gas and do not react chemically.
WARNING: Possible fire hazard. Fuses (i.e., additional sockets) must only be replaced
by fuses of the same type and with the same rating.
WARNING: Possible electric shock hazard. The machine may only be opened by
Aeonmed qualified or Aeonmed authorized service engineer.
WARNING: Electric shock and fire hazard. Do not clean the machine while it is on and
/or plugged in.