RT SensMon: the real-time monitoring value of the inspiratory valve flow sensor, the expiratory
flow sensor, the Airway pressure sensor, the ambient pressure sensor, the oxygen sensor, the
oxygen supply pressure sensor, the Air supply pressure sensor, the breathing system temperature
sensor, and the chassis temperature sensor.
EFM RT Sens Mon: view the real-time monitoring value of the oxygen flow meter sensor, the N
2
O
flow meter sensor, the Air flow meter sensor, and the pressure sensor.
RT Vent Mon: view the real-time monitoring value of the inspiratory volume, the expiratory volume,
the system resistant, the system compliance, and the tube compliance.
SW Versions: the version of GUI, FSU, BDU, PSU, KBD, and EFM.
The Setup sub-menu includes the follow
items.
Language: have several types of language,
for example English, Chinese and so on.
Data: three sections, including
MM/DD/YYYY, DD/MM/YYYY and
YYYY/MM/DD.
Time: two sections, 12 hour or 24 hour
format.
O
2
Sensor: two sections, including Enabled
and Disabled.
Gas Module: two sections, including
Enabled and Disabled.
File Export: enter the password “2020” to
open file export.
Demo: two sections, including OFF and
ON. Click on the OFF button, a correlative
screen appears, see the right picture, enter
the password “2020” to open Demo mode.
5.7.3. Calibration
Oxygen sensor (oxygen cell) or paramagnetic oxygen sensor calibration, gas module calibration,
and touch screen calibration can be achieved through the calibration menu.
Oxygen sensors and paramagnetic oxygen sensors are used to monitor the oxygen concentration
in the gas, one of which is optional for the Aeon8800A Anaesthesia Workstation.