86
Continuous trend information are stored and
shown in the chart, including Pressure,CO
2
,
Agent, MV V
t
, O2 The machine shall
remember the 72 hours trend chart
5.7.5. Tools
The configuration shortcut will become the Tools shortcut without Standby during ventilation. And
the Tools menu contains two sub-menus: PV Tool, RM (Lung recruitment), and CPB.
The Standby mode does not have this function.
When the “Tools” button is selected, the
interface is shown as the figure on the
right.
RM includes sustained inflation and PEEP
increment.
When CPB is turned on, the interface is as
shown in the figure on the right.