9.Technical data
151
9.10. Technical specifications of paramagnetic oxygen module
CAUTION: Specific parameters are subject to the latest technical data released by the
manufacturer.
The oxygen sensor can be used to measure the local oxygen concentration of the Anaesthetic
Workstation. Since it is a consumptive item, the user should pay attention to the validity of the
sensor, and use it according to the features and technical requirements provided by the
manufacturer. The following is main technical parameters of the oxygen sensor for the Aeon8800A
Anaesthetic Workstation.
Table 9-16 Technical specifications of paramagnetic oxygen module
0~100% O
2
(with over range -15% O
2
to 200% O
2
)
Accuracy (Intrinsic Error)
<± 0.4% O
2
in first 24 hours, then <± 0.2% O
2
/week, then <±
0.2% O
2
/month
8~20 seconds dependent on application and filter selection
(biological filter on request)
10kPa-200kPa(1.5psi-30psi)
Within a range of 0℃ to 50℃: Zero<± 0.5% O
2
/10℃, Span: <
±0.5% O
2
/10℃
ROHS Directive 2002/95/EC
Clean dry gas, free of entrained oil, particulates < 3μm,
non-condensing