2.Components
27
2.8. O
2
flush
Press the fast oxygen supply button, 25L/min~75L/min high flow of O2 can be supplied to the fresh
gas outlet.
NOTE: Connect the central gas supply system or the spare gas cylinder and press the
O
2
flush button, high flow O
2
can be delivered to the breathing system and the manual
skin bag, even if the machine is not turned on.
Table 2-10 O
2
flush
2.9. ACGO
ACGO is short for auxiliary common gas outlet.
When the ACGO switch is on, the fresh gas will flow from the ACGO interface to the external
auxiliary manual circuit to provide extra respiratory support to the patients.
Table 2-11 ACGO
CAUTION: ACGO should be firmly connected with external equipment to avoid gas
leakage.
CAUTION: ACGO should be closed to avoid gas leakage when external equipment is
not connected.