8.Regular maintenance
127
8. Regular maintenance
WARNING: To prevent fire:
Please use admitted lubricant for anaesthetic or O
2
device.
Do not use lubricant containing oil or lipin, because when O
2
concentration reaches a certain degree, burning or explosion may
happen.
WARNINGPlease observe disinfection control and safety regulation because the
devices having been used may contain blood and sap.
WARNING: Moving part and removable elements is in danger of jamming hands or
crush. More attention must be paid when moving or replacing system parts.
WARNING: In the process of removing product, strike and quiver must be avoided.
WARNING: Exhausted and old instruments (such as battery and LCD screen) harmful
to environment must be treated according to present local regulations.
8.1. General principle of maintenance
This chapter provides message needed by Aeon8800A Anaesthetic Workstation in term
maintenance. Replace or adjust some components before doing some calibration or adjustment. If
user uses improper components such as improper specification causes malfunction, our company
will not take responsibility. Do not use malfunction device. Replacement and maintenance should
be finished by authorized service engineer or qualified, trained persons with repairmen experience.
After maintenance, the device should be tested and ensure its normal function and comply with
the manufacture’s regulation. Use components manufactured or sold by our company if
replacement is needed, then test and ensure device comply with the manufacture’s specification.
If service support is needed, please call local service engineer. In any case, the repairmen cost will
include the current price and reasonable personnel cost, but except repairmen items within
Aeonmed company’s warranty.
CAUTION: Person without repair experience about such device must not repair this
device.
8.2. Maintenance schedule
Table 8-1 Maintenance schedule
Action
After each repair
Every 12 months
Every 36 months
Check before use
X
X
X
Check before operation
X
X
Check with eyes
X
X