142
Table 9-6 USB interface
Number of USB Interfaces
Standard: 1
Drive Mode
HOST interface, USB1.0/2.0 protocol
Power Supply
5 VDC±5%, 500 mA Max.
Interface Type
USB A-type port
9.6.2. Pneumatic
Table 9-7 Pneumatic
Pipeline supply
O
2
, AIR, N
2
O
Pipeline connection
DISS-male, DISS-female, NIST (ISO 5359)
All fittings used to connect O
2
, N
2
O and Air pipeline gas
supply are already.
APL valve
250kPa
Vaporizer connection
2 Selectable-tecmounts
Pressure display
Color mark list
Pipeline inlet pressure
280600kPa
9.7. Technical specifications of breathing system
Table 9-8 Technical specifications of breathing system
Fresh gas compensation
Flow compensation range: 120L/min
Gases: O
2
, N
2
O, AIR, anaesthetic agent
Absorbent
1500mL(for each absorber canister)
Connection
CGO: ISO5356 connectorstandard 22mm exterior diameter or
15mm interior diameter
Breathing system leak
These data ensures the continuous pressure is higher than the
anticipated pressure during mechanical ventilation.
Pressure: 3kPa, leak volume≤150ml/min
Breathings system
resistance
Use a two-way sine wave test breathing system:
Freq:20/min, Vt=1L,expiration resistence≤0.6kPa inspiration
resistence≤0.6kPa .
Breathing system with small patient resistance should be selected