5.Operating guide
59
5. Operating guide
5.1. Opening
WARNING: Make sure the assembly of the patient breathing tube and controlling
setting are correct before ventilation.
1. Plug the power cord to an electrical outlet.
2. Connect the pipeline gas resource or cylinder gas resource (open the cylinder valve).
3. Turn off O
2
, N
2
O, Air flowmeter and the auxiliary O
2
flowmeter.
4. Turn on the switch.
The user interface lights on, and the
system enters the self-test interface.
5.2. System self–test
The system self-test interface is
displayed.
After the system self-test:
1) Press and confirm “Continue”, the
Automatic Test begins.