58
Set high PAW alarm to 40cmH
2
O.
10. Test low PAW alarm:
Disconnect test lung form breathing system.
Other alarms occur, such as low minute volume alarm.
Make sure low PAW alarm occurs.
11. Test continuous PAW alarm:
Set control items
APL valve:
Adjust to the maximum value
Ventilation Mode:
Manual
Mechanical ventilation stops when the machine operates at Manual ventilation mode.
Block patient end and push O
2
flush button.
Make sure continuous PAW alarm occurs when continuous pressure is at 10cmH
2
O for
15 seconds.
12. Turn off power switch.