4.Pre-operative preparation
57
PAW-High: 40cmH
2
O
Anaesthetic Workstation
All gas flow: OFF
Push O
2
flush to inflate bellows.
4. Make sure:
Initiate mechanical ventilation.
No low gas source pressure alarm occurs.
Ventilator displays correct values.
Bellows moves up and down during mechanical ventilation.
5. Set O
2
flow at 5L/min.
6. Make sure:
End respiratory pressure is within 2cmH
2
O to 3cmH
2
O.
Ventilator displays correct values.
Bellows moves up and down during mechanical ventilation.
7. Test O
2
monitoring and alarm:
Disassemble O
2
sensor, make sure that O
2
takes up 21% of room Air according to the
test result.
Set low O
2
alarm to 50%, make sure low O
2
alarm occurs.
Set low O
2
alarm to 21%, make sure low O
2
alarm is eliminated.
Assemble O
2
sensor in breathing system.
Set high O
2
alarm to 50%.
Push O
2
flush button to inflate bellows.
Make sure high O
2
alarm is initiated.
Set high O
2
alarm to 100%, make sure high O
2
alarm is eliminated.
Put O
2
sensor in pure oxygen; make sure that O
2
takes up 100% according to the test
result.
8. Test low minute volume alarm:
Return to alarm menu.
Set low minute volume alarm limit to 16.0L/min.
Make sure low minute volume alarm occurs.
Return to alarm menu.
Set low minute volume alarm limit to 10.0L/min, make sure low minute volume alarm is
eliminated.
9. Test high PAW alarm:
Set high PAW below peak PAW.
Make sure high PAW alarm occurs.