9.12. EMC Declaration
Below cables information are provided for EMC reference.
Table 9-34 cables information
Cable
Max. cable length,
Shielded/unshielded
Number
Cable
classification
AC cable
5m
Unshielded
1 Set
AC Power
O
2
sensor cable
2.5m
Unshielded
1 Set
DC Power
Mainstream gas monitoring
module cable
2.5m
Unshielded
1 Set
DC Power
Important information regarding Electro Magnetic Compatibility (EMC)
Aeon8800A needs special precautions regarding EMC and put into service according to the EMC
information provided in the user manual; Aeon8800A conforms to this IEC 60601-1-2 standard for
both immunity and emissions. Nevertheless, special precautions need to be observed:
Aeon8800A with Following ESSENTIAL PERFORMANCE is intended used in Professional
healthcare facility environment.
ESSENTIAL PERFORMANCE: IPPV mode V
t
=500, other parameters by default, control is correct.
WARNING: Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of
the Aeon8800A, including cables specified by the manufacturer. Otherwise, degradation of
the performance of this equipment could result.
The use of accessories, and cables other than those specified by AEONMED, with the
exception of accessories and cables sold by AEONMED of Aeon8800A as replacement parts
for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the
Aeon8800A.
WARNING: Use of this equipment Aeon8800A adjacent to or stacked with other equipment
should be avoided because it could result in improper operation.