9.Technical data
169
Table 9-35 EMI Compliance Table
Electromagnetic environment
CISPR 11
Group 1, Class A
Professional healthcare facility environment
Professional healthcare facility environment
Voltage fluctuations
and flicker
Professional healthcare facility environment
NOTE The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) this equipment might not offer adequate protection
to radio-frequency communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.
Table 9-36 EMS Compliance Table
Professional healthcare facility environment
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV Air
3V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fields from
RF wireless
communications
equipment
Rated power
frequency magnetic
fields