9.Technical data
169
Table 9-35 EMI Compliance Table
Phenomenon
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group 1, Class A
Professional healthcare facility environment
Harmonic distortion
IEC 61000-3-2
Class A
Professional healthcare facility environment
Voltage fluctuations
and flicker
IEC 61000-3-3
Compliance
Professional healthcare facility environment
NOTE The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) this equipment might not offer adequate protection
to radio-frequency communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.
Table 9-36 EMS Compliance Table
Phenomenon
Basic EMC standard
Immunity test levels
Professional healthcare facility environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV Air
Radiated RF EM field
IEC 61000-4-3
3V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fields from
RF wireless
communications
equipment
IEC 61000-4-3
Refer to table 3
Rated power
frequency magnetic
fields
IEC 61000-4-8
30A/m
50Hz or 60Hz