9.Technical data
171
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare facility environment
Voltage dips
IEC 61000-4-11
0% U
T
; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and 315º
0% U
T
; 1 cycle and 70% U
T
; 25/30 cycles
Single phase: at 0º
Voltage interruptions
IEC 61000-4-11
0% U
T;
250/300 cycles
Table 9-39 Signal input/output parts Port
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare facility environment
Electrical fast
transients/burst
IEC 61000-4-4
±1 kV
100kHz repetition frequency
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and 80MHz
80%AM at 1kHz
9.13. Essential performance
Oxygen flow under all conditions except the failure of the oxygen supply (pipeline or cylinder)
to the anaesthetic workstation or the generation of a technical alarm condition.
Delivery of a non-hypoxic gas mixture to the patient or generation of a technical alarm
condition.
Non-delivery of excessive concentrations of a volatile anaesthetic agent or generation of a
technical alarm condition.
Airway pressure monitoring and associated alarm.