6.2.2. Alarm message table
Table 6-3
NO
Alarm Name
Priority
Alarm
type
Cause
Remedy
1.
BDU
Communication
failure
High
Technical
alarm
Functional safety
board failure
Start manual ventilation!
Please contact the
manufacturer for
maintenance.
2.
EFM
Communication
failure
High
Technical
alarm
Electronic
flowmeter board
failure
Start manual ventilation!
Please contact the
manufacturer for
maintenance.
3.
FSU
Communication
failure
High
Technical
alarm
Functional safety
board failure
Start manual ventilation!
Please contact the
manufacturer for
maintenance.
4.
Keyboard board
Communication
failure
Low
Technical
alarm
Keyboard board
failure.
Start manual ventilation!
Please contact the
manufacturer for

Table of Contents