6.Alarm and troubleshooting
95
NO
Alarm Name
Priority
Alarm
type
Cause
Remedy
maintenance.
5.
Power module
Communication
failure
Low
Technical
alarm
Power board
failure
Start manual ventilation!
Please contact the
manufacturer for
maintenance.
6.
High oxygen
source pressure
Low
Function
alarm
The oxygen
source pressure
is greater than
680kpa.
Check the oxygen supply.
7.
Low oxygen
supply pressure
High
Function
alarm
The Air source
pressure is lower
than 200KPa
Check the oxygen supply.
8.
Low Air supply
pressure
High
Function
alarm
The oxygen
source pressure
is lower than
200KPa
Check the oxygen supply.
9.
High Airway
pressure
High
Function
alarm
The Airway
pressure
exceeds the
upper alarm limit.
The applied
suction pressure
is higher than the
set value
Check the ventilation
settings.
Check the alarm limits.
The breathing
hose is blocked
or the Airway
intubation is
kinked.
Check the respiratory circuit
and endotracheal intubation.
10.
Low Airway
pressure
Middle
Function
alarm
Insufficient fresh
gas flow,
breathing bag
blocked or
placed
incorrectly
Check the fresh gas
facilities and the position of
the breathing bag.
There is leakage
or falling off
Check the breathing circuit
for closed connection or
leakage.