96
NO
Alarm Name
Priority
Alarm
type
Cause
Remedy
11.
Continuous high
Airway pressure
High
Function
alarm
The Airway
pressure
continued to be
too high.
The patients' autonomous
breathing ability was
examined.
Check the ventilation
settings.
Check the respiratory hose,
respiratory system and
Anaesthesia exhaust
system.
In manual mode, check the
APL valve settings.
The breathing
hose is blocked
or the Airway
intubation is
kinked.
Check the respiratory circuit
and endotracheal intubation.
12.
Unable to reach
the set tidal
volume
High
Function
alarm
There is leakage
or the pipeline
falls off
Check the ventilation
settings.
Check the patient's
compliance.
Check the breathing circuit
for closed connection or
leakage.
Check the endotracheal
intubation
13.
Air loss of
bellows
High
Function
alarm
There is leakage
or the pipeline
falls off
Check the breathing circuit
for closed connection or
leakage
14.
Too much
leakage
High
Function
alarm
There is leakage
or the pipeline
falls off.
Check the breathing circuit
for closed connection or
leakage.
Check the endotracheal
tube or mask.
15.
Apnea
High
Function
alarm
Inability to
breathe or
Start manual ventilation!
Check the ventilation