6.Alarm and troubleshooting
97
NO
Alarm Name
Priority
Alarm
type
Cause
Remedy
ventilate
settings.
The ability of spontaneous
breathing was examined.
16.
Negative
pressure
High
Function
alarm
Insufficient fresh
gas flow,
breathing bag
blocked or
placed
incorrectly.
Check the gas facility and
position of the breathing
bag
Negative
pressure suction
was performed
during
ventilation.
Check the negative
pressure suction device.
Anaesthesia
exhaust system
failure.
Check the function of AGSS.
17.
Apnea backup
ventilation
Middle
Function
alarm
No inspiratory
activity detected
Check the patient's ability of
spontaneous breathing.
Adjust the setting value of
"trigger sensitivity".
Switch to pressure control or
volume control ventilation
mode.
18.
PEEP High
Middle
Function
alarm
The expiratory
branch is
blocked.
Check the breathing hose
and respiratory system at
the expiratory end.
Anaesthesia
exhaust system
failure.
Check the Anaesthesia
exhaust system.
19.
PEEP Low
High
Function
alarm
Insufficient fresh
gas flow,
breathing bag
blocked or
placed
incorrectly.
Check the fresh gas setting
and the position of the
breathing bag.
There is leakage
Check the breathing circuit