98
NO
Alarm Name
Priority
Alarm
type
Cause
Remedy
or the pipeline
falls off.
for closed connection or
leakage.
Anaesthesia
exhaust system
failure.
Check the Anaesthesia
exhaust system.
20.
High tidal
volume
Middle
Technical
alarm
The delivered
inspiratory tidal
volume
exceeded the
upper alarm limit
for 3 consecutive
breathing cycles.
Check the ventilation
settings.
Check the patient's
compliance. Check whether
the patient has self-help
breathing.
Check the alarm limits.
21.
Low tidal
volume
Middle
Technical
alarm
The delivered
inspiratory tidal
volume was
lower than the
lower alarm limit
for 3 consecutive
breathing cycles.
Check the ventilation
settings.
Check the Pmax setting.
Check the patient's
compliance. Check whether
the patient has self-help
breathing.
22.
High minute
ventilation
Middle
Function
alarm
The upper limit of
minute
ventilation alarm
is exceeded.
Check for spontaneous
breathing.
Check the ventilation
settings.
Check the alarm limits.
23.
Low minute
ventilation
Middle
Function
alarm
Minute
ventilation below
alarm lower limit
Check the ventilation
settings.
Check the alarm limits.
Check the endotracheal
tube or mask
24.
High fresh gas
flow
Middle
Function
alarm
The total fresh
gas flow is higher
than 19L/min.
Reduce the flow of fresh
gas.
25.
High respiratory
rate
Middle
Function
alarm
In PS/CPAP
mode, the
Check for spontaneous
breathing.