102
NO
Alarm Name
Priority
Alarm
type
Cause
Remedy
Large dead
volume
Check the ventilation
settings.
46.
CO
2
apnea
High
Function
alarm
Unable to
breathe or
ventilate.
Start manual ventilation!
Check the ventilation
settings.
The ability of spontaneous
breathing was examined.
47.
Various
anaesthetic
gases
At the same
time use
Middle
Function
alarm
A second
anaesthetic was
detected.
Flush the system if
necessary.
Check the fresh gas setting.
Wait for the anaesthetic to
discharge automatically.
48.
Gas module
communication
failed
High
Technical
alarm
There is a
problem with the
connection of the
mainstream gas
module.
Check the connection
between the main gas
module and the Anaesthetic
Workstation.
There is a
problem with the
connection of the
bypass gas
module.
Contact the manufacturer
for maintenance.
The gas module
is damaged.
Contact the manufacturer
for maintenance.
49.
Internal error in
gas module
Low
Technical
alarm
The gas module
is damaged.
Contact the manufacturer
for maintenance.
50.
Gas module
monitoring
concentration
overrun
High
Technical
alarm
The monitoring
value exceeds
the internal limit
of the module.
Check the evaporator
setting.
The gas module
is damaged.
Contact the manufacturer
for maintenance.
51.
Gas module
probe failure
Low
Technical
alarm
Gas module
adapter failure.
Check the gas module
adapter and patient side