6.Alarm and troubleshooting
101
NO
Alarm Name
Priority
Alarm
type
Cause
Remedy
exceeds the
lower alarm limit.
The soda lime
has dried.
Replace sodium lime.
41.
Low
concentration of
inhaled ENF
Middle
Function
alarm
The
concentration of
enflurane inhaled
exceeds the
lower alarm limit.
Check the evaporator and
fresh gas settings.
The soda lime
has dried.
Replace sodium lime.
42.
Low
concentration of
inhaled HAL
Middle
Function
alarm
The
concentration of
inhaled
halothane gas
exceeds the
lower alarm limit.
Check the evaporator and
fresh gas settings.
The soda lime
has dried.
Replace sodium lime.
43.
EtCO
2
High
concentration
Middle
Function
alarm
ETCO
2
has
exceeded the
upper alarm limit.
Check for ventilation.
44.
EtCO
2
Low
concentration
Middle
Function
alarm
ETCO
2
is below
the lower alarm
limit.
Check for ventilation.
45.
High
concentration of
inhaled CO2
Middle
Function
alarm
Sodium lime has
been used up
Check sodium lime.
Add fresh gas.
Check the fresh gas
settings.
The breathing
pipeline leaks
Replace the breathing line.
Inaccurate gas
detection due to
high respiratory
rate.
If necessary, adjust the
alarm limits.