100
NO
Alarm Name
Priority
Alarm
type
Cause
Remedy
for 72 hours
34.
High
concentration of
inhaled N
2
O
High
Function
alarm
The
concentration of
inhaled N
2
O
exceeded 82%.
Check the composition of
fresh gas.
Press the "O
2
+" button
35.
High ISO
concentration by
inhalation
High
Function
alarm
The
concentration of
inhaled
isoflurane gas
exceeds the
upper alarm limit.
Check the evaporator and
fresh gas settings.
36.
High
concentration of
SEV inhaled
High
Function
alarm
The
concentration of
inhaled
sevoflurane gas
exceeds the
upper alarm limit.
Check the evaporator and
fresh gas settings.
37.
High
concentration of
ENF inhaled
High
Function
alarm
The
concentration of
enflurane inhaled
exceeds the
upper alarm limit.
Check the evaporator and
fresh gas settings.
38.
High
concentration of
HAL inhaled
High
Function
alarm
The
concentration of
inhaled
halothane gas
exceeds the
upper alarm limit
Check the evaporator and
fresh gas settings.
39.
Low
concentration of
inhaled ISO
Middle
Function
alarm
The
concentration of
isoflurane gas
inhaled exceeds
the alarm limit.
Check the evaporator and
fresh gas settings.
The soda lime
has dried.
Replacement of sodium lime
40.
Low
concentration of
SEV inhaled
Middle
Function
alarm
The
concentration of
inhaled
sevoflurane gas
Check the evaporator and
fresh gas settings.