7.Cleaning, disinfection, and Sterilization
117
Detach the top plate from the bellows.
Remove the bellows adapter ring from inside the bellows.
Remove the bellows dome U-ring.
NOTE: Dry the bellows by allowing it to hang so that it is fully expanded. This will facilitate
thorough drying and prevent it from sticking to itself.
CAUTION: Do not soak longer than 15 minutes to prevent swelling or aging.
CAUTION: If moisture remains in the bellows after cleaning, it may become tacky.
After all bellows components is completely dry, inspect them for damage before disinfection or