2L/min(Child);
3L/min(Adult)
Have this flow rate triggering function in SIMV~VG、 SIMV~VC、
SIMV~PC、PS/CPAP、BiVent、APRV mode,Error is ±15% or ±1.0L/min,
choose the maximum one
Have this pressure triggering function in SIMV~VG 、 SIMV~VC 、
SIMV~PC、PS/CPAP、BiVent、APRV mode
Have this function in SIMV~VC、SIMV~PC、SIMV~VG、PS/CPAP、
BIVENT、APRV mode,error is ±0.05s or ±20% of the setting , choose
the maximum one
Have this function in APRV mode,error is ±5% or ±20% of the setting ,
choose the maximum one
Have this function in BIVENT、APRV mode,error is ±10% of the setting
or ±3cmH
2
O, choose the maximum one
Have this function in BIVENT、APRV mode,error is ±10% of the setting
or ±3cmH
2
O, choose the maximum one
Have this function in BIVENT、APRV mode,error is ±10% of the setting
or ±0.1s, choose the maximum one
Have this function in BIVENT、APRV mode,error is ±10% of the setting
or ±0.1s, choose the maximum one
10cmH
2
O(Child
); 20cmH
2
O
(Adult)
Have this function in PCV 、SIMV~PC mode,error is ±3cmH
2
O or
±10%, choose the maximum one