146
Item
Range
Stepping
Default Value
Remark
V
SENS
0.5L~15L/min
1L/min
2L/min(Child);
3L/min(Adult)
Have this flow rate triggering function in SIMV~VG SIMV~VC
SIMV~PCPS/CPAPBiVentAPRV mode,Error is ±15% or ±1.0L/min,
choose the maximum one
P
SENS
~20~1 cmH
2
O
1 cmH
2
O
1 cmH
2
O
Have this pressure triggering function in SIMV~VG SIMV~VC
SIMV~PCPS/CPAPBiVentAPRV mode
E
SENS
5~75%
5%
25%
Have this function in SIMV~VCSIMV~PCSIMV~VGPS/CPAP
BIVENTAPRV modeerror is ±0.05s or ±20% of the setting , choose
the maximum one
I
SENS
5~75%
5%
25%
Have this function in APRV modeerror is ±5% or ±20% of the setting ,
choose the maximum one
P
HIGH
5~70 cmH
2
O
1
15(Child);
20(Adult)
Have this function in BIVENTAPRV modeerror is ±10% of the setting
or ±3cmH
2
O, choose the maximum one
P
LOW
3~50 cmH
2
O
1
5
Have this function in BIVENTAPRV modeerror is ±10% of the setting
or ±3cmH
2
O, choose the maximum one
T
HIGH
0.2s30s
Have this function in BIVENTAPRV modeerror is ±10% of the setting
or ±0.1s, choose the maximum one
T
LOW
0.2s30s
Have this function in BIVENTAPRV modeerror is ±10% of the setting
or ±0.1s, choose the maximum one
P
TARGET
5cmH
2
O ~70cmH
2
O
1cmH
2
O
10cmH
2
O(Child
); 20cmH
2
O
Adult
Have this function in PCV SIMV~PC modeerror is ±3cmH
2
O or
±10%, choose the maximum one