9.Technical data
147
Item
Range
Stepping
Default Value
Remark
ΔP
3~60cmH
2
O(SIMV~PC
SIMV~VCSIMV~VG)
OFF 3~60cmH
2
O(PS/CPAP
BiVentAPRV)
1cmH
2
O
5cmH
2
O
Have this function in SIMV~VCSIMV~PCSIMV~VGPS/CPAP
BiVentAPRV mode, error is ±3cmH
2
O or ±10%choose the maximum
one
T
SLOPE
0.2s ~2.0s
0.1sec
0.5sec
error is ±0.5s
P
MAX
10~100cmH
2
O
1mH
2
O
40 cmH
2
O
Have this function in IPPV PCV~VGSIMV~VC mode, error is
±3cmH
2
O or ±10%, choose the maximum one