110
Only the components marked 134 are heat-resistant and
pressure-resistant parts that are capable of withstanding autoclave
sterilization.
All cleaning solutions used must have a pH between 7.0 and 10.5.
CAUTION: Never use the oxygen sensor or its connector in any type of liquid.
CAUTION: Dispose of the oxygen sensor per the manufacturers specification.
CAUTION: Do not use acetic hydro peroxide or formaldehyde steaming.
CAUTION: Do not autoclave the APL valve.
CAUTION: The valve disc is fragile and must, therefore, be handled with care while
removing the valve cage from the valve assembly.
CAUTION: If moisture remains in the bellows after cleaning, it may become tacky.
CAUTION: Do not wash the inner surface of the connector.
CAUTION: Do not autoclave the oxygen sensor.
CAUTION: Prior to use after cleaning or disinfecting, power up the machine and follow
the on-screen prompts to perform the Leak Test and the Compliance Test.
CAUTION: The PAW gauge cannot withstand immersion or the heat and pressure of
autoclaving.
CAUTION: Per acetic acid and formaldehyde fumigation will damage external surfaces.