7.Cleaning, disinfection, and Sterilization
121
Hold the absorber canister and slide it out as
indicated by the arrow, remove it from the
breathing system.
Turn the top cover clockwise or turn the canister
counter-clockwise to open the CO
2
absorber
canister.
Install the CO
2
absorber canister using the reverse
procedure of disassembly.
WARNING: After replacing the fresh
absorbent, please check the airtightness of
the breathing system refer to section
5.1-5.4.
7.4.5.2. Assembly
While noting the previous WARNING, add new pre-pak or loose fill absorbent to the absorber
canisters. Re-install the absorber canisters into the assembly. Rotate the locking mechanism
handle clockwise into the locked position.
Each time the absorbent is replaced, please check the sealing ability of the components by
performing the following operations:
Connect the breathing system and breathing bag to the Anaesthetic Workstation;
Set the Manual/Auto switch to manual state;
Set the APL Valve to 70cmH
2
O;